Best view resolution 1024x768
LIQUID CHROMATOGRAPHY - MASS SPECTROMETER
Model: Waters ZQ-4000
(Installed in 2005)
Model: Waters ZQ-4000
(Installed in 2005)
Applications:
Quantitative analysis, impurity profiling, metabolite studies, pharmacokinetics, pharma and biomedical applications etc.
Specifications:
Mass Range: | 20 - 4000 Da |
Resolution: | 0.2 Da |
Ionisation method: | ESI, APCI |
Analyser: | Single quadrupole |
Detector: | Off-axis Dynolite photomultiplier |
Vacuum system: | Split flow turbo-molecular source pump (160lts/sec) and two stage rotary backing pump |
Computer: | Pentium 4 Processor based Compaq system and Windows NT OS |
Software: | MassLynx 4.0 |
Solvents: | Methanol, Acetonitrile, Dichloromethane, water and chloroform |
Unsuitable solvents: | Solvents like dimethyl formamide (DMF) and dimethyl sulfoxide (DMSO) cannot be used for high viscosity. Default system used here is 50:50 mixture of methanol and water with 0.1% acetic acid added for protonation. |
Instructions to users: | Minimum of 500 µg of sample should be sent for analysis.
|
Contact Tel. No. : 0364 272 1814; 0364 272 1819 ; 0364 272 1810.